Evaluating dental implant stability using three devices Osstell®, Periotest®, and AnyCheck®: a clinical study
1
Al-Falloja Specialized Dental Center, Alanbar Health Directorate, Ministry of Health, Iraq
2
Department of Oral and Maxillofacial surgery, College of Dentistry, University of Baghdad, Bab-Almoadham, Medical City, Baghdad, Iraq
* Correspondence: salwan.bede@gmail.com
Received:
1
October
2023
Accepted:
12
September
2024
Introduction: Implant stability is usually measured with resonance frequency analysis (RFA) and damping capacity assessment (DCA). This study aimed to measure primary and secondary stabilities using 3 devices that are based on these methods, namely; RFA (Osstell®) and DCA (Periotest® and AnyCheck®) to assess the correlations of the measurements obtained by these devices and the correlations between implant stability and insertion torque. Material and Methods: This observational prospective study included 35 dental implants. The implant stability was measured using the 3 devices. Mann–Whitney U test and unpaired t-test assessed the relationship between implant stability and insertion torque, while the Spearman and Pearson correlations measured the correlation between readings collected via the 3 devices for the primary and secondary stabilities. Results: For the primary stability, there was a strong positive correlation between Osstell® and AnyCheck® and moderate negative correlations between Periotest® and both Osstell® and AnyCheck®. While for the secondary stability, strong correlations with similar patterns were observed among the 3 devices. The stability measurements showed significant relationships with the insertion torque. Conclusions: The 3 devices are reliable in measuring implant stability; also, high insertion torque can lead to improved implant stabilities (primary and secondary).
Key words: Implant stability / resonance frequency analysis / damping capacity assessment
© The authors, 2024
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Introduction
The primary stability of dental implants is defined as the lack of clinical movement at the moment of implant insertion, and achieving high primary stability allows for unhindered healing and osseointegration since there is less implant micromotion [1]. In contrast, secondary stability is provided by biological stability via bone remodeling and regeneration [2].
Implant stability can be assessed using various methods, but their dependability was questioned in several studies. The most commonly employed digital techniques are resonance frequency analysis (RFA) and damping capacity assessment (DCA) [3]. The RFA-based device (Osstell®, Goteborg, Sweden) operates predominantly by producing an electric or magnetic impulse that excites a transducer connected to the implant; the resultant oscillation causes a minor implant displacement laterally, with the conversion of the resonance frequency value to implant stability quotient (ISQ), with a greater ISQ score indicating a more stable implant [4].
Periotest® (Medizintechnik Gulden, Germany) is a DCA instrument initially developed to measure tooth movement quantitatively [5]; it is made up of a small computer linked to a handpiece with a tapping rod [6]. Using an accelerometer, the Periotest® calculates the time the tapping head takes to make contact with a tooth or implant. The findings are reported in the form of Periotest values (PTVs), which can range between −8 and +50. Lower numbers indicate higher stability [7].
In recent times, Neobiotech Co., Ltd., Korea developed a modified DCA device (AnyCheck®); like ISQ, the modified DCA device uses an implant stability test (IST) that ranges between 1 and 99, with greater values suggesting a more stable implant [8]. The device impacts the healing abutment six times in three seconds with force smaller than Periotest® and includes an automatic stop feature to preserve the implant when the stability is poor [9]. In clinical terms, an IST of 1–59 suggests poor stability, 60–64 denotes intermediate stability, and >65 represents good stability [10].
Only a few published clinical studies have evaluated the stability of dental implants using the three aforementioned devices, therefore, this study aimed to measure primary and secondary implant stability using 3 devices that are based on RFA (Osstell®) and DCA (Periotest® and AnyCheck®) and to assess the correlations of the measurements obtained by these devices.
Materials and methods
Study design
This prospective observational clinical study was conducted from December 2021 to December 2022. It included patients with single or multiple missing teeth who received dental implants using a delayed implantation protocol.
This study was performed according to the principles of the Declaration of Helsinki (2013) [11], approved by the Research Ethics Committee of the College of Dentistry-University of Baghdad (Protocol number: 412121), and guided by the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) guidelines [12]. After being informed of the nature of the study, the treatment, and the potential complications that may arise, each patient signed an informed consent form.
To ensure meaningful and reliable results the sample size was estimated using G Power 3.1.9.7 for Windows software (Heinrich-Heine University, Dusseldorf, Germany) based on the results of a previous study [1] that investigated the correlation of implant stability values assessed by Periotest and Osstell Mentor devices. The following parameters were used: Tail(s) = Two, Correlation ρ H1 = 0.7 [1], significant level α err prob = 0.05, Power (1-β err prob) = 0.95, Correlation ρ H0 = 0.
Eligibility criteria
The inclusion criteria were normal healthy adult patients with American Society of Anesthesiologists physical status classification category one (ASA I) presenting with healed edentulous areas of adequate bone height and width based on a preoperative CBCT.
The patients were excluded from the study if they demonstrated signs of active infection at the site of the proposed implant, had any uncontrolled systemic disease, or had a history of radiotherapy. Patients who required complicated surgical procedures such as bone augmentation were also excluded from the study.
Radiographic assessment
A preoperative CBCT (Kavo OP 3D PRO, Germany) was performed with the following parameters: 90 Kv, 9.2 mA, 8.1 seconds, (13 × 15) cm field of view, and 0.5 mm slice thickness. The CBCT was used to measure the bone dimensions (height and width) of the proposed implant site and the average bone density using virtual implant properties in the OnDemand3D program (Cybermed Inc., Seoul, Korea). Using the “Implant pick & place tool”, the virtual implant was placed in the proposed implant site (Fig. 1).
Fig. 1 Virtual implant placed in the planned position using OnDemand 3D software. |
Surgical procedure
The dental implants used in this study were T6 standard bone-level implants (NucleOSS™, Izmir, Turkey) with double, reverse buttress, and self-cutting threads, internal connection, and sandblasted and acid-etched (SLA) surface treatment.
All the surgical procedures were carried out under local anesthesia; after reflection of a full-thickness mucoperiosteal flap, the implant site was prepared through sequential drilling (clockwise drilling with a drilling speed of 800 rpm and copious irrigation with normal saline).
The implants were inserted using a motorized technique with a rotational speed of 50 rpm and 50 N/cm torque. A hand ratchet was used to place the dental implant to the desired depth when the insertion torque exceeded 50 N/cm.
After placement of implants, the primary implant stability was measured by the three devices: The first measurement was performed using the RFA Osstell® device and a single-use Smartpeg™ type 21 transducer from Osstell® (Goteborg, Sweden) that is connected to the implant. The measurement was taken from two directions (buccolingual and mesiodistal), the mean of these values represented the primary stability and was recorded as ISQ.
The second measurement was taken by the DCA device Periotest® after connecting a 4 mm height healing abutment to the implant with a standardized torque of 20 N/cm. Three repeated stability measurements were obtained for each implant and the mean stability value was recorded in PTV.
The third measurement was made using the DCA device AnyCheck® and the same healing abutment, two readings were taken for each implant and the mean stability value was recorded as IST.
After measuring the primary stability by the three devices, the healing abutment was removed, the cover screw was applied, and the flap was sutured. Twelve weeks after implant placement, the secondary implant stability was measured similarly using the 3 devices.
Study variables and statistical analysis
The predictor variables were the primary and secondary implant stability measurements using the three devices recorded as ISQ, PTV, and IST, in addition to the insertion torque, which was recorded as ≤50 or >50 N/cm. The primary outcome variable was the correlations of implant stability measurements among the three devices. The secondary outcome variable was the relationship between implant stability and insertion torque.
GraphPad Prism version 6 for Windows (GraphPad Software, La Jolla, CA, USA) was used for the statistical analysis. The descriptive statistics comprised the numbers and the percentages of the categorical variables and the mean, standard deviation (SD), and median of the numerical variables; the distribution of the numerical variables was assessed using the Shapiro-Wilk normality test. The inferential statistics comprised the Mann-Whitney U test, the Unpaired t-test, the Spearman correlation, and the Pearson correlation. Statistical significance was attributed to probability values of ≤0.05.
Results
The study included 16 patients with an age range of 21–60 years and a mean (SD) and median of 46.4 (10.8) and 47.5 years, respectively. They consisted of 11 females (68.8%) and 5 males (31.2%). The patients received 35 implants; the basic clinical characteristics of the dental implants are illustrated in Table I.
For the primary stability, there was a strong positive correlation between Osstell® and AnyCheck® and moderate negative correlations between Periotest® and both Osstell® and AnyCheck®. While for the secondary stability, strong correlations with similar patterns were observed among the 3 devices (Tab. II).
The dental implant inserted with an insertion torque of >50 N/cm demonstrated higher primary and secondary stability measured by the three devices compared to dental implant inserted with ≤50 N/cm insertion torque, as shown in Table III.
The basic clinical characteristics of the dental implants (n = 35).
The correlations between the three devices in primary and secondary dental implant stability measurements.
The relationship between the primary and secondary dental implant stability and insertion torque.
Discussion
Implant stability must be maintained for satisfactory clinical results in dental implants [1]. Evaluating implant stability aids in making appropriate judgments about loading status, allows for good protocol selection on a patient-by-patient basis, identifies conditions when it is preferable not to load, increases patient trust, encourages good communication, and enhances documentation of the case [13]. There are many methods to evaluate implant stability, among which are the RFA and DCA, which are considered non-invasive quantitative objective methods [14].
In the current study, the mean values of the secondary stability measured by the 3 devices were slightly lower than the mean values of the primary stability; this finding is in line with Gomez-polo et al., who suggested that high primary implant stability tends to decrease with time; in contrast, low primary implant stability tends to increase over time [15]. In addition, Andersson et al. maintained that the decrease in primary implant stability over time could be due to mechanical relaxation and/or remodeling of the bone in response to high stresses encountered during dental implant insertion [16]. Nevertheless, both primary and secondary stability values recorded in this study are considered to be high according to the standards of the used devices that consider ISQ values >70 using Osstell®, PTV of −8 to 0 using Periotest®, [14] and IST values >65 using AnyCheck® to indicate high stability [8].
The primary and secondary stability measurements showed moderate to strong correlations among the three devices, which concurs with Pyo et al., who demonstrated that high IST values were associated with high ISQ values and low PTV values [17]. Also, Oh et al. showed that an increase in ISQ value is accompanied by a decrease in PTV value at different points of time after installation of the dental implant and vice versa and concluded that the findings obtained from Periotest and Osstell Mentor were correlated [7].
The PTV is inversely related to implant stability, with low values indicating good stability and high values indicating poor stability. PTVs between −8 and 0 indicate efficient osseointegration and “good stability,” while values between 1 and 9 indicate “moderate stability”, and values between 10 and 50 indicate “poor stability” [14]. The 3 devices detected the high primary and secondary stabilities recorded in this study with moderate to strong correlations, suggesting that they were reliable in measuring implant stability.
A recent study by Shim et al. (2023), on the other hand, demonstrated a general pattern of weak to moderate correlations in stability measurements among the three devices at different time points, the authors reported a strong negative correlation between PTV and IST values only in one follow up visit at 2 months after surgery [18].
The present study showed that increasing insertion torque leads to high primary and secondary implant stability, which is in keeping with Joshi et al., who maintained that the primary and secondary implant stabilities are predominantly influenced by insertion torque [19]. On the other hand, Cassetta et al. demonstrated that implants installed with an insertion torque >50 N/cm do not necessarily lead to higher secondary stability. As a result, increasing insertion torque during implant installation may result in enhanced primary implant stability without enhancing secondary implant stability [20]. This may be due to the fact that high insertion torque may lead to high compression forces that may cause local microcirculation disturbance, necrosis of the osteocyte, and bone resorption [21]. Also, Lages et al. reported in a systematic review that included 12 studies that the operators should define just one method for the assessment of dental implant stability because RFA and insertion torque are independent and non-comparable techniques for the evaluation of primary implant stability [22]. Noaman et al., however, showed a non-significant correlation between insertion torque and implant stability [23]. Moreover, Degidi et al., in their study that included a sample of 4135 dental implants, concluded that insertion torque and ISQ value appear to be independent features of primary implant stability, and only insertion torque is affected by bone density [24].
It is noteworthy to mention that implant stability measurement with DCA devices is associated with some technical limitations. Obtaining valid readings requires that the patient be seated in an upright position and the device be held in a horizontal posture with an acceptable upward or downward angulation of about 25 degrees, which may result in slight difficulty in measuring stability in posterior implants, also a certain distance should be maintained between the tip of the probe and the implant during measurement. Lee et al. (2023) in an in vitro study demonstrated that the reliability of Periotest and Anycheck (DCA devices) was significantly affected by the position of the bone model and implant location and that Osstell showed consistently reliable results [25].
The results of this study need to be interpreted after considering its main limitation related to the observational design.
Within the limitations of this study, it can be concluded that RFA (Osstell®) and DCA (Periotest® and AnyCheck®) are reliable in measuring implant stability and that high insertion torque can improve primary and secondary implant stability.
Funding
This article received no specific funding.
Conflicts of interest
The authors declare that they have no conflict of interest.
Data availability statement
The data will be provided upon request.
Ethics approval
This study was conducted in accordance with the Declaration of Helsinki guidelines and was approved by the institutional Research Ethics Committee (protocol # 412121).
Informed consent
Each patient signed informed consent to participate in this study.
References
- Oh JS, Kim SG. Clinical study of the relationship between implant stability measurements using Periotest and Osstell mentor and bone quality assessment. Oral Surg Oral Med Oral Pathol Oral Radiol 2012;113:e35-e40. [PubMed] [Google Scholar]
- Elsawy MK, Eldestawy MT, Madany UM. Assessment of Osstell and Periotest systems in measuring immediate dental implants stability. Al-Azhar J Dent Sci 2020;23:173–179. [CrossRef] [Google Scholar]
- Al-Samman AA, Al-Rawee RY, Tawfeeq BA. Does primary stability is mandatory for dental implant success? a systematic review of literature. J Dent Oral Sci 2022;4:1–16. [Google Scholar]
- Buyukguclu G, Ozkurt-Kayahan Z, Kazazoglu E. Reliability of the Osstell implant stability quotient and penguin resonance frequency analysis to evaluate implant stability. Implant Dent 2018;27:429–433. [CrossRef] [PubMed] [Google Scholar]
- Di Stefano DA, Arosio P, Capparè P, Barbon S, Gherlone EF. Stability of dental implants and thickness of cortical bone: clinical research and future perspectives. a systematic review. Materials 2021;14:7183. [CrossRef] [PubMed] [Google Scholar]
- Kocak-Buyukdere A, Kayabay G. Measurement of dental implant stability by two different techniques. Niger J Clin Pract 2021;24:973. [CrossRef] [PubMed] [Google Scholar]
- Oh JS, Kim SG, Lim SC, Ong JL. A comparative study of two noninvasive techniques to evaluate implant stability: periotest and Osstell mentor. Oral Surg Oral Med Oral Pathol Oral Radiol 2009;107:513–518. [CrossRef] [Google Scholar]
- Lee J, Pyo SW, Cho HJ, An JS, Lee JH, Koo KT, et al. Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study. J Periodontal Res 2020;50:56–66. [Google Scholar]
- Lee DH, Shin YH, Park JH, Shim JS, Shin SW, Lee JY. The reliability of anycheck device related to healing abutment diameter. J Adv Prosthodont 2020;12:83. [CrossRef] [PubMed] [Google Scholar]
- Kim YW, Min SK, Wang JW, Shin KS. Assessment of factors that affect primary stability of minimally invasive implant (MII): an in vitro study. J Implantol Appl Sci 2022;26:1–14. [CrossRef] [Google Scholar]
- World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA 2013;310:2191–2194. [CrossRef] [PubMed] [Google Scholar]
- Vandenbroucke JP, von Elm E, Altman DG, Gotzsche PC, Mulrow CD, Pocock SJ, et al. Strobe Initiative, Strengthening the Reporting of Observational Studies in Epidemiology (STROBE): explanation and elaboration. Int J Surg 2014;12: 1500–1524. [CrossRef] [PubMed] [Google Scholar]
- Mistry G, Shetty O, Shetty S, Singh R. Measuring implant stability: a review of different methods. J Dent Implant. 2014;4:165. [CrossRef] [Google Scholar]
- Andreotti AM, Goiato MC, Nobrega AS, da Silva EVF, Filho HG, Pellizzer EP. et al., Relationship between implant stability measurements obtained by two different devices: a systematic review. J Periodontol 2017;88:281–288. [CrossRef] [PubMed] [Google Scholar]
- Gómez-Polo M, Ortega R, Gómez-Polo C, Martin C, Celemin A, del Rio J. Does length, diameter, or bone quality affect primary and secondary stability in self-tapping dental implants?. J Oral Maxillofac Surg 2016;74:1344–1353. [CrossRef] [PubMed] [Google Scholar]
- Andersson P, Verrocchi D, Viinamäki R, Sennerby L. A one-year clinical, radiographic and RFA study of neoss implants used in two-stage procedures. Appl Osseointegration Res 2008;6:23–26. [Google Scholar]
- Pyo SW, Kim HG, Kwon O, Otgonbold J, Lee KW. Reliability verification of damping capacity assessment through in vitro analysis of implant micromotion in peri-implant bone loss model. Int J Oral Maxillofac Implants 2021;36:106–114. [CrossRef] [PubMed] [Google Scholar]
- Shim JS, Kim MY, An SJ, Kang ES, Choi YS. Evaluation of implant stability according to implant placement site and duration in elderly patients: a prospective multi-center cohort study. J Clin Med 2023;12:5087. [CrossRef] [PubMed] [Google Scholar]
- Joshi S, Dhatrak P, Nimbalkar S, Gherde C. An effect of various parameters on insertion torque to improve the success rate of dental implantation: a review. Mater Today: Proc 2021;43:928–934. [CrossRef] [Google Scholar]
- Cassetta M, Brandetti G, Altieri F. Are the insertion torque value and implant stability quotient correlated, and if so, can insertion torque values predict secondary implant stability? A prospective parallel cohort study. Int J Oral Maxillofac Implants 2022;37:135–142. [CrossRef] [PubMed] [Google Scholar]
- O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res 2000;2:85–92. [CrossRef] [PubMed] [Google Scholar]
- Lages FS, Douglas-de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: a systematic review. Clin Implant Dent Relat Res 2018;20:26–33. [CrossRef] [PubMed] [Google Scholar]
- Noaman AT, Bede SY. The relationship of implant stability quotient and insertion torque in dental implant stability. J Baghdad Coll Dent 2022;34:29–35. [CrossRef] [Google Scholar]
- Degidi M, Daprile G, Piattelli A. Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants. Clin Implant Dent Relat Res 2012;14:501–507. [CrossRef] [PubMed] [Google Scholar]
- Lee HN, Kim MS, Lee JY, Xu Zihan, Ryu JJ, Shim JS. Reliability of implant stability measuring devices depending on various clinical conditions: an in vitro study. J Adv Prosthodont. 2023;15:126–6. [CrossRef] [PubMed] [Google Scholar]
Cite this article as: Dhahi AY, Bede SY. 2024. Evaluating dental implant stability using three devices Osstell®, Periotest®, and AnyCheck®: a clinical study. J Oral Med Oral Surg. 30: 20.
All Tables
The correlations between the three devices in primary and secondary dental implant stability measurements.
The relationship between the primary and secondary dental implant stability and insertion torque.
All Figures
Fig. 1 Virtual implant placed in the planned position using OnDemand 3D software. |
|
In the text |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.