The relationship between the primary and secondary dental implant stability and insertion torque.

Insertion torque N/cm Osstell/ISQ
Periotest/PTV
AnyCheck/IST
Mean SD Median Mean SD Median Mean SD Median
Primary stability measurements
≤50 72.9 3.6 72.0 –3.2 1.9 –2.6 73.3 4.9 74.0
>50 76.9 3.8 77.0 –5.1 1.5 –4.8 76.5 4.1 77.0
P-value 0.002 [S] 0.002* [S] 0.035 [S]

Secondary stability measurements
≤50 70.1 7.8 74.2 –1.4 3.5 –2.5 72.1 4.1 72.5
>50 76.2 4.7 76.5 –4.3 1.4 –4.3 76.1 3.0 76.0
P-value 0.022 [S] 0.003* [S] 0.002* [S]

ISQ, Implant stability quotient; PTV, Periotest value; IST, Implant stability test; SD, Standard deviation; S, significant.

Mann Whitney U test.

*

Unpaired t-test.